Chamber THERMASTER

Temperature Control During Sun Protection Testing

The THERMASTER is specifically engineered to control and ensure substrate surface temperature during in vitro solar testing, adhering to various standards. It is fully compatible with PMMA substrates molded HD6 and sandblasted SB6.

The metal support, with the substrate on top, maintains temperature by inertia during the weighing and application stages, which take less than 2 minutes.

The THERMASTER has been specifically designed to control and guarantee substrate surface temperature during the in vitro solar testing process, in order to comply with the various standards.
Fully suitable for PMMA substrates molded HD6 and sandblasted SB6 the metal support (with the substrate on top) is used to maintain temperature by inertia during the weighing and application stages (less than 2 minutes).

 

Directions of use

The THERMASTER developed by WENEOS is an easy-to-use thermal enclosure that maintains the substrate surface temperature throughout an in vitro solar test:

  • Switch on the unit and set the desired temperature
  • Position PMMA substrates molded (HD6) or sandblasted (SB6) on the metal supports
  • The metal support with the substrate resting on it is used to maintain temperature by inertia during the weighing and spreading steps using the robot SPREADMASTER robot (usually in under 2 minutes)
  • The substrate is placed back on the THERMASTER (in the dark) to ensure surface temperature control during the drying stage

Weneos THERMASTER – Specifications

Technical InformationTHERMASTER Specifications
DimensionsW: 300 mm x D: 210 mm x H: 120 mm
Weight2.3 kg
Termperature• Accuracy: 0.1 °C
• Operating range: Ambient – 45°C (safety limit at 50°C max)
Location9 locations with metal gondolas
Compliance• ISO 23675:2024 – SPF in vitro
• ISO 23698:2024 – HDRS
• ISO 24443:2021 – UVAPF in vitro
• FDA monograph 2011 – Broad Spectrum
• Boots Star Rating system 2011 – UVA/UVB ratio
Security• 2014/30/EU Electromagnetic Compatibility Directive (EMC),
• 2011/65/EU Restriction of the use of certain hazardous substances (RoHS),
• 2011/35/EU Low-voltage Directive,
• Harminized standards (EN 61000-6-1, EN 61000-6-2, EN 61000-6-3, EN
61000-6-4, EN 62061)
Power100-240 VAC / 50-60 Hz
IP classIP42
Condition of use• Temperature : [5° – 45°C]
• Humidity: [10% – 80%] non-condensing