Labsphere UV3000S Sunscreen Tester

Labsphere’s UV3000S Sunscreen Tester incorporates the latest technologies and software enhancements to achieve accurate in vitro SPF Sun Protection Factor analysis of cosmetics and suncare products to qualify and receive the “very high, 50 or higher SPF” sun protection label, and HEVIS sun protection testing. Driven by industry requirements to simplify product labeling and improved in vitro methods to validate product UV Protection, the UV3000S is designed to comply with approved in vitro methods and in vivo spectrophotometer methods.

  • Extended test capabilities covering UV SPF and Hyperpigmentation HEVIS
  • Improved calibration validation efficiencies
  • Repeatable sample placement
  • Automated sample stage
  • Versatile testing options

The UV3000S flash technology saves time and money when bringing new products to market. The UV3000 rapidly measures the diffuse transmittance of sunscreen samples in the ultraviolet wavelength region from 290 nm – 450 nm and High Energy Visible (HEV) blue light from 400 nm to 500 nm. Labsphere’s Spectralon® integrating sphere incorporates a re-optimized xenon flash lamp to provide exceptional diffuse illumination of the product sample and minimize data integration time. Our new highperformance diode array spectrometers coupled by advanced fiber optics are optimized at the system level for low stray light with superior wavelength stability and flash-to-flash repeatability.

Stay current with the latest global cosmetic testing and reporting of SPF, UVA, UVB, UVR, and more:

  • ISO23675
  • ISO 23698
  • ISO24443
  • FDA 2012
  • COLIPA
  • Boots Star Rating

UV3000S Sunscreen SPF Tester: Key Benefits

  • NEW robust industrial design – built to last!
  • Save time and money with Labsphere’s flash technology for rapid testing – perform a full sample study in minutes!
  • NEW low learning curve menu-driven software for sample testing and reporting to industry standards plus new configurable interface and reporting features
  • NEW know more with extended test capabilities over the UV-VIS wavelength range from 290 to 500 nm for testing UV SPF as well as Hyperpigmentation HEVIS
  • NEW save time and trust results with our calibration validation kit with linear filter holder that improves calibration validation efficiencies by more than 75%
  • Achieve repeatable sample placement with the manually operated sample stage positioning for pre- and post-irradiation
  • NEW be more efficient with our optional Automated Sample Stage
  • Versatile testing options with 50mm, 75mm and ISO 9 point sample test templates for repeatable testing and reporting

Wavelength Range

290 to 500 nm*

Wavelength Accuracy

±1 nm

Bandwidth

(FWHM) < 4 nm

Wavelength Step

(Data Interval) 1 nm

Optical Geometry

Hemispherical Illumination/0° viewing (d/0)

Integrating Sphere Material

Spectralon®

Integrating Sphere Port Area

< 5%

Sample Exposure Area

0.79 cm2

Lamp

Xenon Flash Lamp

UV Dose Per Measurement Cycle

< 0.2 J/cm2

Sample Positioning Stage

Manual Stage

Measurement Range

Transmittance: 0 – 100%

Absorbance: 0 – 3.0 A (Dual Doped PMMA Method)

SPF: 1 – 50+

Scan Time

< 5 s

Computer Interface

USB

Power Requirements

110 – 120/220 – 240 VAC, 60/50 Hz

Operating Environment

00 – 50°C, 0% – 70% RH (non-condensing)